Safety risks and 'blank screen' in defibrillators prompt SA ambos warning
Resumen detallado
The South Australian government will provide ambulances with back-up defibrillators after paramedics are warned about potential safety issues to its current equipment, including a delay in defibrillating patients.
South Australian ambulances will be fitted with back-up defibrillators after safety issues have been flagged regarding current equipment.
The ABC has seen an SA Ambulance Service internal document that warns staff about the risks of defibrillators currently fitted in ambulances.
The federal government said although a suspension remains in place on the supply of the defibrillators, existing devices can continue to be used.
The South Australian government will provide ambulances with back-up defibrillators after paramedics are warned about potential safety issues to its current equipment, including a delay in defibrillating patients.
The ABC has seen a "safety alert" issued by SA Ambulance Service (SAAS) to staff this morning about defibrillators, known as Corpuls 3 monitors, currently fitted in ambulances.
In the document, paramedics were warned the device's monitor may show a blank screen and stop displaying patient information for 5 to 25 seconds.
Other risks highlighted to staff included "a planned shock not being delivered when the shock button has been pressed firmly" and an inability to defibrillate patients for up to 15 seconds.
Staff were also told the device could report "incorrect heart rate values" and a system to automatically recognise a STEMI — a threatening type of heart attack — may fail.
Another risk identified was the unit "unexpectedly rebooting without warning".
Inicio del artículo, extracto de ABC Business. Para leer el resto, abra el artículo en el sitio del medio.
MarketNews muestra el titular y el resumen publicados por el medio. El artículo completo se lee en el sitio del editor. (www.abc.net.au)